F7.2 Candela Fluid High temperature

F7.2 Candela Fluid High temperature

How to fix F7.2 fault message on Candela pro series devices First, we need to understand the causes of this fault F7.2 which is mainly because Distilled or DI Water is over temperature. Now to the fix, how do we fix this problem easily?There’s a couple of...
F7.3 Candela Fluid Flow

F7.3 Candela Fluid Flow

How to fix F7.3 fault message on Candela pro series devices First, we need to understand the causes of this fault F7.3 which has several reasons: Distilled or DI water pump pressure fault Low or no Distilled or DI water pressure and/or flow. Distilled or DI water...
F7.4 Candela Fluid Temperature Sensor

F7.4 Candela Fluid Temperature Sensor

How to fix F7.4 fault message on Candela pro series devices First, we need to understand the causes of this fault F7.4 which is because of: Temperature sensor fault (sensor circuit open or shorted) Now to the fix, how do we fix this problem easily?There’s a...
F8.2 Candela DCD Canister High Pressure

F8.2 Candela DCD Canister High Pressure

How to fix F8.2 fault message on Candela pro series devices First, we need to understand the causes of this fault F8.2 which is mainly because Faulty pressure sensor, or overheated DCD canister. Now to the fix, how do we fix this problem easily? Check DCD canister...
F8.3 Candela Handpiece DCD Valve.

F8.3 Candela Handpiece DCD Valve.

How to fix F8.3 fault message on Candela pro series devices First, we need to understand the causes of this fault F8.2 which is mainly because of the DCD valve fault.. Now to the fix, how do we fix this problem easily? Replace the delivery system If problem persists,...