F5.2 Candela Power Supply Charge

F5.2 Candela Power Supply Charge

First, we need to understand the causes of this fault F5.2 which is mainly because of the high voltage power supply charge time-out. Now to the fix, how do we fix this problem easily?There’s mainly two solutions for this problem: Turn off the laser system for 30...
F5.4 Candela Power Supply Charging Rate

F5.4 Candela Power Supply Charging Rate

How to fix F5.4 fault message on Candela pro series devices First, we need to understand the causes of this fault F5.4 which is mainly because of the high voltage power supply is set to incorrect charge rate. Now to the fix, how do we fix this problem...
F6.1 Candela Calibration Problem

F6.1 Candela Calibration Problem

How to fix F6.1 fault message on Candela pro series devices First, we need to understand the causes of this fault F6.1 which is mainly because the laser system failed to find threshold HD energy within normal voltage or laser system failed to find max HD energy within...
F6.5 Candela Calport Switch

F6.5 Candela Calport Switch

How to fix F6.5 fault message on Candela pro series devices First, we need to understand the causes of this fault F6.5 which is mainly because of the Calport redundant switches indicate different states for longer than 1 second. Now to the fix, how do we fix this...
F7.1 Candela Fluid low temperature

F7.1 Candela Fluid low temperature

How to fix F7.1 fault message on Candela pro series devices First, we need to understand the causes of this fault F7.1 which is mainly because Distilled or DI Water is under temperature. Now to the fix, how do we fix this problem easily?There’s a couple of...